Industrial Laser Monitoring System
High Resolution Focus Imaging – High Damage Threshold (Up to 1* kW),USB 3.0
The ILMS is specially designed for profiling focused, high-power industrial lasers. This system combines reimaging/magnification optics, a polarization preserving beam sampler, and a DataRay beam profiler to measure small beam waists which would otherwise damage a traditional profiling system. Magnification of the focused beam allows full pixel-by-pixel 2D measurements of beam spots as small as a few microns.
The ILMS is compatible with most DataRay profilers and supported by the full-featured, highly customizable, and user-centric software (included without licensing fees). The software automatically accounts for the magnification of the system, so results do not require post processing or corrections.
Features
Options from 190nm up to 16 μm
Beam waist diameters down to a few μm
High magnification options available (50X and beyond)
Capable of measuring high power beams, up to kW range possible
Three swappable filters for flexible attenuation
Profiler easily removed from system for stand-alone use
Optional calibrated pinhole aperture denotes working distance
and helps provide protection against damage from beam misalignment
Available without beam sampler for focused, low power beams
Integrated power meter options (contact us for more information)
Applications
Tightly focused beams, fiber ends, edge couplers, laser diodes, and more
High power laser cutting systems
Telecommunications
Quality control
* Supported beam diameter and power limits are model-dependent. Please contact us to discuss options for custom requirements.
PLEASE NOTE:
This product requires 64-bit Windows 7, 8/8.1, 10, or 11
Specification Detail Magnification
ILMS-5-X: 5x
ILMS-10-X: 10x
ILMS-50-X-: 50xAR Coating Wavelength (nm)
ILMS-X-UV: 250-425
ILMS-X-VIS: 425-675
ILMS-X-532: 532
ILMS-X-NIR: 750-1550
ILMS-X-1064: 1064Input Optics NA
ILMS-5-X: 0.17
ILMS-10-X: 0.17
ILMS-50-532-: 0.6
ILMS-50-1064: 0.65System Dimensions (mm)
ILMS-5-X: 70.00 x 133.25 x 176.30
ILMS-10-X: 70.00 x 133.25 x 252.50
ILMS-50-X: 70.00 x 133.25 x 328.70Typical minimum beam waist diameter (1/e², um)
ILMS-5-X: 11
ILMS-10-UV: 5.5
ILMS-10-VIS: 5.5
ILMS-10-NIR: 9
ILMS-50-532: 2
ILMS-50-1064: 3Integrated Beam Profiling Camera
ILMS-X-UV: WinCamD-LCM
ILMS-X-VIS: WinCamD-LCM
ILMS-X-532: WinCamD-LCM
ILMS-X-NIR: WinCamD-LCM
ILMS-X-1064: WinCamD-LCMPixel Count, H x V
WinCamD-LCM: 4.2 MPixel, 2048 x 2048
Optical Format
WinCamD-LCM: 1"
Image Area
WinCamD-LCM: 11.3 x 11.3 mm
Pixel Dimension
WinCamD-LCM: 5.5 x 5.5 µm
Shutter Type
WinCamD-LCM: Global
Max. Frame Rate*
WincamD-LCM: 60 Hz
Single Pulse Capture PRR
WinCamD-LCM: USB 2.0 @ 6.3 kHz; USB 3.0 @ 12.6 kHz
Signal to RMS Noise
WinCamD-LCM: 2,500:1 (34 dB optical / 68 dB electrical)
Electronic Shutter Range
WinCamD-LCM: USB 2.0 @ 12,600:1 (41 dB), USB 3.0 @ 25,000:1 (44 dB)
ADC
WinCamD-LCM: 12-bit
Measurable Sources
WinCamD-LCM: CW beams, pulsed sources; CW to 12.6 kHz with single pulse isolation
Measured Beam Powers
Model dependent, contact sales@dataray.com
Manual Beam Attenuation
Includes ND-1, ND-2, and ND-4 filter inserts
Displayed Profiles
Line, 2D & 3D plots. Normalized or un-normalized.
Linear or Logarithmic, Zoom x10
2D, 3D in 10, 32 or max. colors or grayscale
Contoured display at 10 and 32 colorsMeasured & Displayed Profile Parameters
Raw and smoothed profiles
Triangular running average filter up to 10% FWHMBeam Diameter
Diameter at two user set Clip levels
Gaussian & ISO 11146 Second Moment beam diameters
Equivalent diameter above a user defined Clip level
Equivalent Slit and Knife Edge diametersBeam Fits
Gaussian & Top Hat profile fit & % fit
Equivalent Slit profileBeam Ellipticity
Major, Minor & Mean diameters. Auto-orientation of axes.
Centroid Position
Relative and absolute
Intensity Weighted Centroid and Geometric Center
Beam Wander Display and StatisticsMeasurement Accuracy (not limited to pixel size)
0.1/M µm processing resolution for interpolated diameters (where M is magnification)
Absolute accuracy is beam profile dependent - ~1/M µm accuracy is frequently achievable (where M is magnification)
Centroid accuracy is also beam dependent (as good as ±1/M µ since it is arithmetically determined from all pixels above the centroid clip level) (where M is magnification).Processing Options
Image & profile Averaging, 1, 5, 10, 20, Continuous.
Background Capture and Subtraction.
User set rectangular Capture Block for capture
User set or Auto ellipse Inclusion region with beam tracking for processing
*.ojf files save all WinCamD custom settings for particular test configurationsPass/Fail display
On-screen selectable Pass/Fail colors. Ideal for QA & Production.
Log Data and Statistics
Min., Max., Mean, Standard Deviation, to 4096 samples
Relative Power Measurement
Rolling histogram based on user's initial input. Units of mW, µJ, dBm, % or user choice (relative to a reference measurement input)
Fluence
Fluence, within user defined area
Certification
RoHS, WEEE, CE
Multiple Cameras
Up to 4 cameras, parallel capture.
1 to 8 cameras, serial capture.Optical depth from housing/filter to sensor (no window) ±0.2 mm
4.0/ 8.5 mm (verify before purchasing)
Mounting
M6 and M3 mounting points
Weight‚ Camera w/ MagND and filter cover
5 lbs (2kgs)
Minimum PC Requirements
Windows 7/8/8.1/10/11 64-bit, 4 GB RAM, USB 2.0/3.0 port
* Capture block size dependent
商品属性 [波长] 190nm up to 16 μm [光束尺寸] down to a few μm