DP SERIES POCKELS CELL DRIVERS FOR LOW OPERATION VOLTAGES (≤2 MHZ, ≤3.6 KV)
DP series Pockels cell drivers feature HV pulse amplitude up to 3.6 kV and HV pulse repetition rate up to 2 MHz. OEM version available upon request.
Features:
HV pulse amplitude up to 3.6 kV
Repetition rate up to 2 MHz
HV pulse duration extension using pulse regeneration technique
Alternative aluminum case option featuring conductive cooling possibility and relocated water and HV output connectors (Option 1)
OEM version available upon request
DP series units are available in two versions: "open frame" which is ideal for OEM manufacturers incorporating drivers in their own systems and encased in aluminum housings. Encasing of Pockels cell driver in aluminum housing solves two problems: shields both humans and electronics from high voltage impact from operating Pockels cell driver, and protects driver itself from potentially harmful external contact - ensuring safe operation and driver longevity. The housed option is especially handy for researchers and custom product manufacturers who use these drivers during their own systems build-up.
DP-50-3.6
DP-50-3.6-AIDP-250-2.5
DP-250-2.5-AlDP-250-3.6
DP-250-3.6-AlDP-400-1.5
DP-400-1.5-AIMaximal HV operating voltage
3.6 kV
2.5 kV
3.6 kV
1.5 kV
Maximal HV repetition rate
60 kHz
250 kHz
250 kHz
400 kHz
Pulse duration
100 - 5000 ns
HV pulse rise time, typical
<7 ns
<6 ns
<7 ns
<5.5 ns
HV pulse fall time, typical
<7 ns
<6 ns
<7 ns
<5.5 ns
Output polarity
Positive
HV power consumption
<20 W
<40 W
<75 W
<20 W
12V / 24V power consumption
<1 W
<4.5 W
<4 W
<5.5 W
Cooling
Water
DP-500-2.6
DP-500-2.6-AlDP-1000-1.8
DP-1000-1.8-AlDP-2000-1.5
DP-2000-1.5-AlMaximal HV operating voltage
2.6 kV
1.8 kV
1.5 kV
Maximal HV repetition rate
500 kHz
1 MHz
2 MHz
Pulse duration
100 - 5000 ns
HV pulse rise time, typical
<6.5 ns
<6 ns
<7 ns
HV pulse fall time, typical
<6.5 ns
<7 ns
<7 ns
Output polarity
Positive
HV power consumption
<90 W
<80 W
<120 W
12V / 24V power consumption
<6 W
<9 W
<12 W
Cooling
Water
Driver needs to be mounted on heatsink unless it is cooled by water. Driver's base plate temperature needs to be lower than 35 °C in all regimes of operation. Power consumption for 6 pF load.
Voltage or repetition rate derating is necessary if capacitance of your Pockels cell is higher. Contact vendor for details.
商品属性 [HV工作电压] 2.5 kV,Max [HV重复频率] 250 kHz,Max [脉冲持续时间] 100 - 5000 ns [输出极性] Positive [HV功耗] <40W