FAST RESPONSE LASER SENSORS
DESIGNED FOR OEM AND LABORATORY APPLICATIONS
OEM Amplified BLINK FR
A SUPERIOR TECHNOLOGY
OEM Amplified BLINK FR sensors are based on a proprietary technology which enables natural response times of 70 ms (typ.) while still keeping broadband spectral range, power density capability and direct high power operation up to 60W of comparable thermopile detectors.
The response time is a key parameter for laser detectors where a faster response is required because several phenomena like pointing stability, fast drifts, instabilities can only be monitored if the detector responds very quickly. However, when the detector is fast it cannot withstand high powers (as in the case of all photodetectors) or, vice-versa, when the detector can withstand the power, it is slow (as it happens with thermopile sensors or Peltier devices).
Photodetectors can respond to pulses < 1ns. They can only withstand few milliwatts of direct laser power and their advantage of a fast response is, in most of applications, spoiled by standard readout electronics which curb the response time to 0.1-0.2 s.
Thermopile detectors are inherently slow: their natural response times range from > 1s to several tens of seconds, depending on the maximum power rating. Those natural response times can be reduced by electronics circuits and speed-up algorithms, but cannot be lower than 800ms – 1s, and only in the case of very low power heads.
OEM Amplified BLINK FR sensors, differently from thermopile sensors, offer an unreached natural (without any acceleration by additional electronics) response time down to 70 ms (typ.). This is a game changing feature for laser power measurement because it paves the way to a number of applications that were not possible before.
TECHNICAL SPECIFICATIONS
POWER MODE
Max. Average Power
50 W
Min. power
50 mW
Response time (0-90%)
Typ. 70ms (min. 50 ms – max. 90 ms)
Power Calibration Uncertainty
± 3%
Power linearity
± 5%
ABSORBER SPECS
Aperture
16 mm x 16 mm
Type
K
Absorber Spectral Range
0.2 - 25 µm
Calibration Spectral Range
0.25 - 1.1 µm ; 10.6 µm
Max Power Density (1)
1.5 KW/cm²
Max Energy Density J/cm² (2)
1 J/cm²
AMPLIFIER SPECS
Amplifier Input Voltage
±7 to ±12 VDC,
or 14 to 24 VDC floatingOutput Voltage @ Full Scale
5 V
Min Detectable Voltage
5 mV
Sensitivity
100 mV/W
GENERAL CHARACTERISTICS
Cooling
Water (a)
Weight
300 g
Head dimensions
56 x 56 x 34 mm
Cable length - connector
1.5 m - Bare Wire
NOTES
(1) Measured at 1064 nm, 10 W. Damage thresholds depend on power level.
(2) 10ns @ 1064 nm(a) Water min. 2 l/min, max. 4 l/min (@ 15 - 30 °C); admissible rate of water temperature variation < 1 °C/min
商品属性 [波长范围] 0.2 – 25 µm [功率范围] 50 mW - 50W