Mini TPA/PD
Compact Size Autocorrelator for Routine Tasks
Mini is the perfect combination of tuning-free autocorrelation measurement, compact size and high sensitivity.
Compact DesignWith its compact footprint of only 160 x 220 x 155 mm, the Mini autocorrelator is perfect for working environments where space is limited. When it is needed elsewhere, simply place this portable unit in the lightweight aluminum case provided, for easy and safe transportation to the next site.
NIST Traceable Calibration
Laboratories and manufacturers are often faced with systematically establishing an unbroken chain of calibrations to specified references. All APE autocorrelator models are calibrated to a traceable standard in accordance with NIST (U.S. National Institute of Standards and Technology) measurement traceability specifications. A printed and signed calibration certificate is provided with each pulseCheck.
Tuning-Free Wavelength Matching with Mini TPA
Conventionally, autocorrelators used to split an optical pulse into two replicas and recombine them for the Second Harmonic Generation (SHG) in a nonlinear crystal. The APE Mini TPA instead benefits from the two photon absorption principle. This eliminates the need of SHG crystal angle tuning and makes the wavelength tuning process unnecessary.
Switching Between Collinear and Non-Collinear Mode with Mini PD
The Mini PD supports quick and easy switching between collinear and non-collinear measurement modes. While collinear mode provides pulse width measurement with additional qualitative information about the chirp, non-collinear mode gives background-free autocorrelation with a high dynamic range.
UV at 250 nm to IR at 3200 nm
APE provides a selection of exchangeable Optics Sets, ranging from UV at 250 nm to IR at 3200 nm, for sensitive measurements across an extremely broad wavelength region. Due to its compact footprint, the Mini TPA is also the perfect answer to your space-saving and easy portability requirements.
Mini Autocorrelator Specification Mini TPA Mini PD Measurable Pulse Width Range
50 fs ... 3.5 ps;
Optional: down to 35 fs50 fs ... 3.5 ps;
Optional: down to 10 fsAvailable Optics Sets
Exchangable Optics sets of TPA-type
Fixed Optics set of PD-type - not exchangable
Wavelength Range
UV 250 ... 400 nm;
VIS 1 400 ... 700 nm;
NIR 700 ... 1250 nm;
IR 1250 ... 2100 nm;
Ext. IR 2000 ... 3200 nmVIS 1 420 ... 550 nm;
VIS 2 540 ... 750 nm;
NIR 700 ... 1200 nm;
IR 1100 ... 1800 nm
(Fixed range to be specified on order)Type of Measurement Mode
Collinear intensity
Non-collinear intensity, collinear interferometric
Measurement Mode Switching
No
Yes
Sensitivity*
Depending on Optics set,
typically 0.1 W2;
UV range < 500 W2Depending on Optics set,
typically 1 W^2SHG Tuning for Phase Matching
Not required
Software aided
Max Input Power
300 mW or 5 µJ (whichever results in lower value)
1 W (e.g. oscillator with a rep. rate of approx. 70 MHz) or 10 µJ (e.g. amplifed
Recommended Repetition Rate
> 300 Hz;
For UV only: upper limit 2 MHz> 300 Hz
Delay Resolution
< 0.001 % of scan range
< 0.001 % of scan range
Delay Linearity
< 1 %
< 1 %
Trigger Mode
300 Hz ... 50 kHz
300 Hz ... 50 kHz
Input Aperture
6 mm (free-space)
6 mm (free-space)
Input Beam Coupling
Free-space; Optional: fiber coupling FC/PC, FC/APC, SMA
Free-space; Optional: fiber coupling FC/PC, FC/APC, SMA
Input Polarization
Linear horizontal, vertical available as option
Linear horizontal, vertical available as option
Beam Input Height
86 ... 150 mm; Optional 50 mm
86 ... 150 mm; Optional 50 mm
Software
Included; Real-time display of pulse width and central wavelength, different fitting routines
Included; Real-time display of pulse width and central wavelength, different fitting routines
Fitting Routine
Gaussian, Sech2, Lorentz
Gaussian, Sech2, Lorentz
Connection
USB
USB
Remote Control
Possible via TCP/IP (SCPI command set)
Possible via TCP/IP (SCPI command set)
Calibration
NIST traceable calibration certificate included
NIST traceable calibration certificate included
* Measured sensitivity including Optics Set, defined as average power times peak power of the incident pulses P(AV) * P(peak)