MPM-210/MPM211,212,213
Multi-channel Optical Power Meter for Swept Test System
MPM-210H/MPM-211, 212, 213, 215
The perfect tool for very fast simultaneous measurement of multi-port optical devices
High Speed Measurement
(Up to 1 million logging sampling points per port)Up to 20-Port Measurement
Power Meter Module (MPM-211 / 212 / 215)
Current Meter Module (MPM-213)
Our MPM-210H is ideal for IL, WDL and PDL measurement of multi-port optical components including Dense Wavelength Division Multiplexing (DWDM), Arrayed Waveguide Gratings(AWG) , Wavelength Selective Switches (WSS) and more. When combined with our TSL-Series laser equipped with a power monitor output, the MPM-210 allows the user to complete real-time high precision IL measurements.
High speed measurement
Up to 1 million logging sampling points per port 2 memory buffers per port
Fast measurement (100 kHz) with high resolutionHigh dynamic range
- 80 to + 10 dBm (MPM-211 / 212)
- 70 to + 5 dBmA (MPM-215)High dynamic range (logging mode)
50 dB per scan (MPM-211 / 212)
70 dB per scan (MPM-215)Current meter module (MPM-213)
Dynamic current range - 70 to + 10 dBmAAnalog output (MPM-212)
Up to 20-port measurement
Wavelength Range 1250 to 1650 nm
Main frame, MPM-210H
Parameter
Unit
Specifications
Notes
Module number
–
Up to 5
Interface
For Power meter
–
GP-IB, Ethernet, RS-232C
For System
–
USB
Trigger input
–
TTL (3.3 V)
BNC
Trigger output
–
TTL (3.3 V)
BNC
Power monitor
V
0 to 3
BNC
Supply voltage
V
AC 100 to 240, 50 / 60 Hz
Maximum power consumption
VA
50
Operating temperature
℃
10 to 40
Operating humidity
%
< 80
non condensing
Weight
kg
6
Dimensions (W) x (D) x (H)
mm
210 x 350 x 133
Power meter module, MPM-211(4ch), MPM-212(2ch), MPM-215(4ch)
Parameter
Unit
MPM-211
MPM-212
MPM-215 *2
Notes
Sensor element
-
InGaAs
Wavelength range
nm
1250 to 1680
Specification wavelength range
nm
1250 to 1630
Power dynamic range
dBm
- 80 to + 10
- 70 to + 5
Dynamic range @Fixed gain (typ.) *2, 3
dB
45
70
Logging mode
Number of gain range
-
5
1
Maximum safe power
dBm
+ 16
Total uncertainty
%
+/- 5 @- 60 to 9 dBm
+/- 5 @>-55 dBm
Power resolution
dB
0.001
Linearity *2, 4
dB
+/- 0.03 @- 55 to 9 dBm
+/- 0.02 @>-40 dBm
+/- 0.05 @>-50 dBmPolarization dependent responsivity (typ.) *2, 5
dB
< 0.025
@1525 to 1585 nm
< 0.03
@1270 to 1630 nm
Averaging time
sec
10μ to 10
Data logging capability
–
1,000,000 x 2 buffer per port
Port number per module
–
4 ports
2 ports
4 ports
Analog output *6
None
With
None
Connector type
–
FC
Current meter module, MPM-213
Parameter
Unit
Specifications
Notes
Current dynamic range
dBmA
- 70 to + 100@ 100pA to 10 mA
Dynamic range @Fixed gain (typ.) *2, 3
dB
45 dB
Logging mode
Number of gain range
-
4
Maximum safe power
dBmA
16
Total uncertainty
%
+/- 5 @- 45 to 9 dBm
Averaging time > 10 ms
+/- 1 @>- 35 dBm (typ.)
Averaging time > 10 ms
Power resolution
dB
0.001
Linearity
dB
+/- 0.03@ -45 to 9 dBmA
Averaging time > 10 ms
Averaging time
sec
10μ to 10
Data logging capability
–
4 ports
Port number per module
–
1,000,000 per port x2
Connector type
–
BNC Connecter
Reverse Bias Voltage
V
-–
*1: 1: This module is not compatible with other modules on a same Main Frame.*2: Temperature: 23 +/- 5 ℃, *3: Averaging time >50us, Within ±1 ºC after zeroing, At gain 1, 2 or 3 for MPM-211/212 *4: For MPM-211/212, Averaging time >100ms, Auto range mode, For MPM-215, Averaging time >30us *5: SMF, Straight connector
*6: Output voltage 0 to 2 V, Measurement dynamic range > 30 dB (typ.)Common information for module
Parameter
Unit
Specifications
Notes
Operating temperature
℃
10 to 40
Operating humidity
%
<80
non condensing
Dimensions (W) x (D) x (H)
mm–
30.3 x 183.5 x 114.8
Connectors are not included.
Typical swept test system configuration
IL/PDL measurement setup with the polarization controller PCU-110 and the power meter MPM-210H
IL measurement setup with the power meter MPM-210H
Ordering Code