TPIR-785 Raman
Highest-Sensitivity Raman Spectrometer
TPIR-785 Configurations
TPIR-785 Raman systems come with a variety of accessories to fit your research and application needs.
» Sampling accessories for a broad range of sample formats
» Switching between accessories is quick and easy
» Flexible adaptor and filter chamber for customized setup
Standard Configuration
785 nm Raman probe
» Point-and-shoot operation
» Choice of FC/PC or SMA connectors to
choose from
» Round-to-linear collection fiber bundle
for best efficiency
» Universal fiber adaptor can accept
different fiber inputs (including customized
probes)
Routine Sample
Measurement
785 nm Raman cuvette holder
» Based on proprietary CUBE accessories
» Pre-aligned CUBE stack has built-in 785 nm
Raman filters and focusing optics
» Plug and play without the need for alignment
» Ideal for routine sample measurement
Filter Chamber
Configuration
Extended functionality
» Most flexible setup for different spectroscopic
measurements
» Filter chamber has built-in focusing and
collimation optics and accepts standard 25 mm
diameter filters
» Expands system functionality beyond 785 nm
Raman
» Universal fiber adaptor can accept different
fiber inputs
SPECIFICATIONS
Spectrograph
Aperture ratio
f/2 Spectral resolution
5 cm-1 (High Resolution*);7.8 cm-1 (Extended Coverage**) Spectral range
80 – 2250 cm-1 (High Resolution*);80 – 3650 cm-1 (Extended Coverage**) Total system throughput
>68% Laser*** Temperature-stabilized
785 nm multi-mode laser; 475 mW (software adjustable) Integration time
10 µs – hours *with 1200 g/mm grating and 10 µm slit width
**with 830 g/mm grating and 10 µm slit width
***contact your sales engineer for other excitation laser wavelengths
Detectors
Standard(PIXIS 100/400 BR EX4)
Scientific CCD with eXcelon™ 4 technology;
Eliminates fringes and improves QE in the NIR range;
1340 x 100/400 pixel array with 20 x 20 µm pixel size;
-80°C cooling for low dark current
High QE(BLAZE® 100/400 HR)
Proprietary scientific CCD delivers superior NIR QE;
1340 x 100/400 pixel array with 20 x 20 µm pixel size;
-95°C cooling for ultralow dark current
Accessories / System
785 nm Raman probe
Stainless steel probe head with anodized aluminum body;10 mm focal length;Round-to-linear fiber bundle for high light-collection efficiency
Universal fiber adaptor
Accommodates SMA, FC/PC, or 10 mm ferrules
Slit
Manual adjustable slit (10 µm – 3 mm)
IntelliCal®
Automated wavelength and intensity calibration (optional)
CUBES
CUBES accessories for other Raman and spectroscopic setups (optional)
Raman notch filter chamber
Flexible filter chamber for Raman setups (optional)
Software
LightField® for system control and data collection; SDKs in Python® (Python Software Foundation), LabVIEW® (National Instruments), and MATLAB® (MathWorks)
Connectivity
USB 3.0
Operating temperature
+5°C to +30°C
Storage temperature
-20°C to +50°C
Size
24" x 17 3/8" x 7 3/8" (61 cm x 44 cm x 19 cm)
Weight
60 lbs. (27 kg)
商品属性 [波长范围] 80 – 2250 cm-1 /80 – 3650 cm-1 [光学分辨率] 5 cm-1 /7.8 cm-1