PIXIS-XO Soft X-Ray Cameras
The PIXIS-XO series of fully integrated imaging cameras utilizes back illuminated and back illuminated deep depletion CCDs without AR coating, for direct detection of the widest range of X-rays between ~ 30 eV and 20 keV (AR coated devices are not useful for X-ray energies < 500eV). With a 1340 x 400 imaging array, 20 µm pixels,
100% fill factor, low noise electronics and -90° C thermoelectric cooling with either air or water, this system is ideal for worry-free operation in research and OEM environments. The rotatable conflat flange with high-vacuum-seal design, software selectable gains and readout speeds make these cameras well suited for ultra-high vacuum applications.
FEATURES
BENEFITS
Back-illuminated deep depletion and Back-illuminated
CCD, with no AR coating
Provides very low X-ray flux imaging, high sensitivity and high spatial resolution
2 Mhz / 16-bit readout
100 kHz / 16-bit readout
High speed readout for rapid image acquisition; Slow speed readout for high sensitivity
with wide dynamic range, high signal-to-noise ratio (SNR) and excellent energy resolution
Software selectable gains for each digitization speed
Allows optimization of system performance for lowest noise to highest SNR
1340 x 400 image area,
20 x 20 µm pixels
Spectroscopy format designed for high frame rate imaging
Ultra low noise electronics
Best possible system performance
Flexible user-selectable binning & readout
Total flexibility to optimize experiments and SNR
Kinetics
Custom readout mode offers microsecond resolution
Deep thermoelectric air cooling
Maintenance-free operation - NO need for a liquid circulator or additional power supply
Deep thermoelectric water cooling
Vibration free operation
Conflat vacuum interface
Industry-standard, high-vacuum compatibility
TTL input and output
External Trigger input with programmable polarity;
TTL output with exposure or readout monitor
USB 2.0 interface
Seamless, plug-and-play connection to PC notebooks & desktops; Easy OEM integration
WinView/Spec software (for Windows XP/7; 32-bit)
or LightField® (for Windows 7; 64-bit)
Powerful, yet easy-to-use software packages for acquisition, display and analysis;
Powerful yet intuitive interface; Direct stream to hard drive
PVCAM/PICAM interface
Universal programming interface for easy custom programming; Compatible with
Windows XP/7 (32-bit), Windows 7 (64-bit) and Linux; PICAM for Windows 7 (64-bit)
LabView™ Scientific Imaging Tool Kit (SITK®)
Predefined vis for easy integration of camera controls into large experiment
SPECIFICATIONS
PIXIS-XO: 400B
PIXIS-XO: 400BR
Features
Back-illuminated CCD without AR coating, for
sensitivity between ~30 eV to 10 keV
Back-illuminated, deep depletion CCD without AR
coating, for sensitivity for widest range of X-ray
detection between ~ 30eV to 20 keV.
CCD Image Sensor
Princeton Instruments exclusive; scientific grade 1;
MPP; back-illuminated; no AR coating
Princeton Instruments exclusive; scientific grade 1;
NIMO; back-illuminated deep depletion;
no AR coating
Dark current @ -75° C
(with ambient air @ +20° C)
0.001 e-/p/sec (typical)
0.005 e-/p/sec (max)
0.03 e-/p/sec (typical)
0.065 e-/p/sec (max)
CCD format
1340 x 400 imaging pixels; 20 x 20 mm pixels; 100% fill factor; 13.3 x 13.3 mm (optically centered)
Deepest cooling temperature,
TE air cooling1
(with ambient air @ +20° C)
-90° C typical; -75° C guaranteed
Thermostating precision
±0.05° C
Cooling method
Thermoelectric air or liquid cooling (CoolCUBE II required)
Full well
Single pixel:
100 ke- (typical), 60 ke- (minimum)
High Sensitivity node: 250 ke- (typical), 220 ke- (minimum)
High Capacity node: 1000 ke- (typical), 750 ke- (minimum)
ADC speed/bits
100 kHz/16-bit and 2 MHz/16-bit
System read noise @100 kHz
@2 MHz
3.0 e- rms (typical),
5 e- rms (max)
11 e- rms (typical), 16 e- rms (max)
Vertical shift speed
<15 msec/row (programmable)
Non-linearity
<1% @ 100 kHz
<2% @ 2 MHz
Software selectable gains
1, 2, 4 e- (high sensitivity); 4, 8, 16 e- (high capacity); available at all speeds
Operating systems supported
Windows XP/Vista/7; Linux
Data interface
USB2.0 (5m interface cable provided); Optional Fiberoptic interface is available for remote operation
I/O signals
Two MCX connectors for programmable frame readout, shutter, trigger in
Operating environment
+5° C to +30° C non-condensing
Bakeout temperature
70° C (maximum)
Vacuum Compatibility
10 -8 Torr
Certification
CE
Dimensions / Weight
16.59 cm (6.53") x 11.81 cm (4.65") x 11.38 cm (4.48") (L x W x H) / 2.27 kg (5 lbs)
NOTES: All specifications subject to change
1. The minimum temperature attainable is dependent on the vacuum condition - temperature can be lowered w/lower vacuum
Spectral Rates
@100 kHz Full Vertical Binning (FVB)
60 fps*
@ 2 MHzFull Vertical Binning (FVB)
315 fps
@ 2 MHz(0.2 mm high)
1300 fps
* frames per second
Quantum Efficiency Curve
4.5" Conflat
6" Conflat
商品属性 [CCD Image Sensor] Princeton Instruments exclusive; scientific grade 1; NIMO; back-illuminated deep depletion; no AR coating [CCD format] 1340 x 400 imaging pixels; 20 x 20 mm pixels; 100% fill factor; [Dark current] 0.03 e-/p/sec (typical) 0.065 e-/p/sec (max) [Imaging area] 13.3 x 13.3 mm (optically centered)