Terahertz Spectroscopic System TAS7400
Low-cost, multi-purpose terahertz spectroscopic system to meet diverse applications
Key Features
The TAS7400 product line offers a series of low-cost, all-purpose spectroscopic systems that enable spectroscopic measurements using terahertz (THz) waves. The tools are capable of performing non-destructive analysis on a wide variety of sample types, making it applicable to a broad range of fields from life sciences to electronics, where precise chemical and material characterization is critical. The systems are ideal for settings ranging from basic research to product development, as well as for manufacturing and quality control, and employ easy-to-use spectroscopic methods that do not require specialized knowledge of THz wave generation or optics.
Low-cost, all-purpose THz spectroscopic system offers superior spectroscopic performance
Compact desktop form factor
THz spectroscopic analysis in a range of frequencies (0.03–7 THz)
Transmission, reflectance, ATR (Attenuated Total Reflection), and transmission polarization measurement modes support spectroscopic analysis of a wide variety of materials
Dry air filter unit attached, so that measurements will be unaffected by water vapor absorption
A full range of holders and cells to handle aqueous solutions, powders, and other types of samples
Superior Performance
High Troughput
Advantest’s proprietary sampling
technique—an electronically
controlled sweep method—delivers
higher throughput than any previous
system.
Highly stable measurement reproducibility
Advantest’s proprietary low-jitter optical fiber
laser technology and independently developed
analog signal analysis technology enables highly
stable reproduction of refractive index (phase)
measure-ments to within ±0.03 % (typ.).
Wideband THz spectroscopic analysis
The TAS7400SP, along with two specialized
spectroscopic analysis systems—the
TAS7400SL and the TAS7400SU—provide
bandwidth coverage required to serve a
broad array of applications.
Key Specifications
Terahertz Spectroscopic System
TAS7400SL
TAS7400SP
TAS7400SU
(low-frequency system)
(standard system)
(broadband system)
Primary measurement
applications
Spectroscopic analysis
(transmission, reflectance)*1
Spectroscopic analysis
(transmission, reflectance,
ATR, transmission polarization
analysis)*1
Spectroscopic analysis
(transmission, reflectance, ATR)*1
Analytical object
Dielectric / chemical materials, others
Chemical / industrial / biological materials, pharmaceuticals, others
Specimen dimensions Transmission /
reflectance modes
φ20 mm ~ 30 mm, < 10 mm thick φ5 mm ~ 30 mm, < 10 mm thick ATR mode — < φ5 mm (powders, liquids), φ5 mm ~20 mm, < 10 mm thick (solids) Transmission /
polarization analysis mode
— φ5 mm ~ 30 mm, < 10 mm thick Performance Frequency range*2
0.03 ~ 2 THz
0.1 ~ 4 THz
0.5 ~ 7THz (transmission / reflectance modes)
0.5 ~ 6.5THz (ATR mode)
Frequency accuracy*2
< ±10GHz at 0.56THz
< ±10GHz at 1.4THz
< ±10GHz at 1.4THz
Frequency resolution 1.9GHz / 7.6GHz
1.9GHz / 7.6GHz
1.9GHz / 7.6GHz
Dynamic range*2
(at peak frequency)
> 50dB
> 60dB
> 57dB (transmission / reflectance modes)
> 55dB(ATR mode)
Throughput
200msec / scan
Measurement accessories
Transmission mode / transmission polarization analysis mode (SP only):solid sample holder,
liquid / powder cells*3, dry air purge kit*3, revolving holder*3
Reflectance mode:reflectance mirror, revolving holder*3
ATR mode(SP/SU only):powder holder
Display
Spectral display(transmittance, reflectance, ATR*, phase difference, absorbance, absorption
coefficient, complex refractive index, complex permittivity
*ATR supports only SP/SU
Time response display (electric field strength)
Software*3
Transmission polarization analysis application, automatic control, FDA21CFR Part11 support,
offline analysis
Dry air purge
External dry air unit(external air supply necessary)
External accessory*4
Thermal control accessory(2 models available: -10 ~ +80℃ and room temperature ~ +300℃)
Controller
Standard OS:Windows7 Pro. 64bit
Data file formats
Binary format, JCAMP-DX, SPC, CSV
General specifications
Operating temperature range:+10 ℃ ~ +30℃, relative humidity: <80% (with no condensation)
Storage temperature range:-10℃ ~ +50℃, relative humidity: <80% (with no condensation)
Analysis unit power source:AC100V(100-120)/ 200V( 220- 240)±10%, 50/60Hz, 180VA
Measurement unit power source:AC100V(100- 120)/ 200V( 220-240)±10%, 50/60Hz,
150VA (excluding PC)
Analysis unit dimensions: 430(W)× 540(D)× 330(H)mm, weight: <30kg
Measurement unit dimensions:500(W)×490( D)× 410(H)mm, weight: <40kg
*1:When purchasing a terahertz spectroscopic system, users must select at least one measurement accessory.
*2:At temperatures of 23℃±5℃
*3:Option
*4:Option for transmission accessory only
Thermal Control Accessory Specification
TAS1020
TAS1030
Notes
Temperature range
-10.0 ~+80.0℃
Room temperature ~ +300℃
— Resolution
0.1℃
1.0℃
— Control interface
USB
Can be controlled independently of system
Accessories supported
Transmission accessory
—