FTIR Process Gas Analyzer
The Gasmet FCX-Series features a comprehensive range of high performance FTIR process gas analyzers. The wall mounted Gasmet FCX process gas analyzers offer reliable results even when analyzing complex gas mixtures.
The Gasmet FCX process gas analyzers incorporates a Fourier Transform Infrared spectrometer, gas cell, and signal processing electronics, a built-in Industrial Computer and a TFT display into a compact enclosure.
The analysis results of Gasmet FCX process gas analyzer are transmitted via 4-20 mA analog output signals or digital outputs (ModBus)
The enclosure has IP 65 protection and the analyser is equipped with Vortex cooling system.
The sample cell can be equipped with an optional heating and temperature control up to 180 °C. Sample cell absorption path length is selected according to the application.
General parameters
Measuring principle: Fourier transform infrared, FTIR
Number of components: 1 - 50 components
Response time, T90: < 120 s, depending on the gas flow and
measurement time
Operating temperature: 5 – 30 °C, non-condensing
Climate control: Vortex
Storage temperature: -20 - 60°C, non-condensing
Power supply: 100-115 or 230 V / 50-60 Hz
Power consumption: 500 W
Spectrometer
Resolution:8 cm-1 or 4 cm-1
Scan frequency:10 or 4 scans / s
IR source:SiC, 1550 K
Detector:Peltier cooled MCT
Beamsplitter:ZnSe
Window material:BaF2 or ZnSe
Wave number range:900 - 4200 cm-1
Sample cell
Structure: Multi-pass or single pass cells. Available from 1 cm up to 9.8 m
Standard materials: Stainless Steel or 100 % gold or rhodium coated
aluminium
Connectors: Inlet Swagelok 6 mm Outlet Swagelok 8 mm
Optional ¼”
Sample cell gaskets: Viton®. Kalrez® available upon special request
Temperature: 180 °C, maximum
Window material: BaF2 or ZnSe or AR coated ZnSe
Measuring parameters
Zero point calibration: 24 hours, calibration with nitrogen
(5.0 or higher N2 recommended)
Zero point drift: < 2 % of measuring range per zero point calibration interval
Sensitivity drift: None
Linearity deviation: < 2 % of measuring range
Temperature drifts: < 2 % of measuring range per 10K temperature change
Pressure influence: 1 % change of measuring value for 1 % sample
pressure change.Ambient pressure changes measured and compensated
Gas inlet and outlet conditions
Gas temperature: Non-condensing, the sample gas temperature should
be the same as the sample cell temperature
Flow rate: 120 - 600 liters per hour
Gas filtration: Filtration of particulates (2 µm)required
Sample gas pressure: Ambient
I/O
I/O is located in FCX in stand-alone version. When FCX is delivered
with Gasmet MSS or MSSH sampling system, I/O is located in the
sampling system.
Standard:
• 4 X change-over relay output. 125VAC/0.5A 30VDC/2A max.
• 1,2,3, or 8 analog outputs 4 - 20mA
• Digital output terminal, 8 digital outputs 24 VDC, 0.5A
• Digital input terminal, 8 digital inputs 24 VDC, single-conductor
connection, 3.0ms input filter
Options:
• 4 analog inputs 4 – 20mA
• Fieldbus (ModBus, Profibus)
Computer
Processor: 2 GHz Intel Celeron
Memory: 256 MB SDRAM
Hard disk: > 40 GB
Operating system: Windows 7 (32-bit)
Application software: Calcmet for Windows
Outputs: Isolated 4-20 mA for each component (maximum is 16) or
digital RS-232 Modbus
Relays: Sample gas; Function alarm;Service request; Concentration
alarm.
Display: Door mounted 12” TFT display
Air conditioning
Gasmet FCX is equipped with Vortex cooling system. Other cooling systems
are available on request.
Cooling capacity: 270 W (air 7 bars, 21 ºC, dewpoint –40 ºC)
Air consumption: 425 Nl/min, maximum
Air quality: Pressure 7-10 bars
Temperature + 43 ºC, maximum Dewpoint –40 ºC
Options
Sample pump:For ambient air up to 50 °C only
Automatic zero valve:Up to 50 °C sample temperatures
Remote control:Modem and software for remote diagnostics
Sample cell temp.:50 - 180 °C
Fan cooling:Removes Vortex, no IP rating
Pressure measurement:Inside sample cell
Enclosure
Material:Stainless steel
Dimensions (mm):800 * 600 * 300
Weight:50 kg
Protection:IP 65
CE label:According to EMI guideline 89/336/EC
Electronics
A/D converter: Dynamic range 95 dB
Signal processor: 32-bit floating point DSP 120 MFLOPS