BeamGage Training DVD
SP90429
How to become a great user of the world’s most powerful laser beam analysis software https://youtu.be/w9JkKIXcF5s
SP300 Beam Profiling Camera
The SP300 camera accurately captures and analyzes wavelengths from 190nm - 1100nm. It features a compact design, wide dynamic range, unparalleled signal to noise ratio, and reduced blooming that makes it ideal for measuring CW and pulsed laser profiles.
1928x1448 pixel resolution with a 3.69µm pixel pitch
USB 3.0 for faster data capture rates
BeamGage Standard or Professional software included
Item
Specification
Model
SP300
Application
1/1.8” format
Spectral Response
190 - 1100nm (2)
Active Area
7.1mm x 5.3mm
Pixel spacing
3.69μm
Number of effective pixels
1928 x 1448
Minimum system dynamic range
56 dB
Linearity with Power
±1%
Accuracy of beam width
±2%
Frame rates in 12 bit mode (4)
26 fps at full resolution
Shutter duration
30μs to multiple frames
Gain control
0 dB to 24 dB
Trigger
Hardware/Software trigger & strobeout
Photodiode trigger
Si response: SP90408
Saturation intensity (1)
0.97µW/cm2
Lowest measurable signal (1)
1.2nW/cm2
Damage threshold
50W/cm2 / 0.1J/cm2 with all filters installed for < 100ns pulse width(3)
Dimensions
44 mm x 29 mm x 58 mm
CCD recess
17.5 mm
Image quality at 1064nm
Pulsed with trigger sync - excellent
Pulsed with video trigger - good
CW - good
Operation mode
Double tap interline transfer CCD
Software supported
BeamGage STD or PRO
PC interface
USB 3.0
OS Supported
Windows 7 (64) and Windows 10
Notes:
(1) Camera set to full resolution at maximum frame rate and exposure times, running CW at 632.8nm wavelength. Camera set to minimum useful gain for saturation test and maximum useful gain for lowest signal test.
(2) Camera may be useable for wavelengths below 350nm but sensitivity is low and detector deterioration may occur. Therefore UV image converter is recommended. Although our silicon cameras have shown response out to 1320nm it can cause significant blooming which could lead to significant errors of beam width measurement. We would suggest our XC130 InGaAs camera for these wavelengths to give the best measurements.
(3) This is the damage threshold of the filter glass of the filters. Assuming all filters mounted with ND1 (red housing) filter in the front. Distortion of the beam may occur with average power densities as low as 5W/cm2.
(4) Highly dependent on PC processor and graphics adapter performance.
Drawing
Ordering
BGS-USB3-SP300
SP90375
BeamGage Standard software, software license, 1/1.8” format 1928x1448 pixel camera with 17.5mm C mount CCD recess. Comes with USB 3.0 cable and 3 ND filters
BGP-USB3-SP300
SP90376
BeamGage Professional software, software license, 1/1.8” format 1928x1448 pixel camera with 17.5mm C mount CCD recess. Comes with USB 3.0 cable and 3 ND filter
BGS TO BGP UPGRADE
SP90233
Upgrade BeamGage Standard Edition to Professional Edition. Requires a new camera key to activate.
BeamGage Training DVD
SP90429
How to become a great user of the world’s most powerful laser beam analysis software https://youtu.be/w9JkKIXcF5s
商品属性 [波长] 190 - 1100nm [光束尺寸] 37μm - 5.3mm [兼容的光源] CW, Pulsed [传感器] Silicon CCD