iDus 401 Series
The Scientific Grade, 1024 x 127 deep depletion spectroscopy CCD camera is ideally suited to rapid analysis, multi-channel and low-light applications including fluorescence and Raman spectroscopy. The 401’s range of Front and Back Illuminated and Deep Depletion sensor options provide optimised performance from UV through Visible to Near IR.
Peak QE of 95%
TE cooling to -100°C
UltraVac™
Single window design
Fringe suppression technology
Overview
The USB2.0-based iDus series is a compact, yet feature-rich platform suitable for demanding spectroscopy applications such as low-light UV/NIR Photoluminence or Raman spectroscopy, as well as day-to-day routine laboratory operation and integration into industry-grade systems.
The iDus 401 boasts market-leading deep thermo-electric cooling to -100°C, complemented by Andor’s UltravacTM technology with its un-matched reliability track record in the scientific and industrial community. The iDus low-noise electronics allows seamless access to the best signal-to-noise ratio performance for the widest range of photon detection regimes.
Anti-fringing comes as standard on back-thinned & back-illuminated sensors for this 1024 x 127, 26 μm pixels array optimized for high dynamic range, high resolution and rapid data acquisition from Far UV to NIR.
Specifications
Key Specifications Model No.
DV401A
DU401A
DU401A BR-DD
Sensor Options
BVF: Back Illuminated CCD, Vis-optimized and anti-fringing
FI: Front Illuminated CCD
BR-DD: Back Illuminated CCD, Deep Depletion with anti-fringing optimum sensor for Near IR applications
Active Pixels
1024 x 127
Pixel Size
26 x 26 µm
Image Area
26.6 x 3.3 mm with 100% fill factor
Minimum Temperature Air Cooled
-55 °C
-80 °C
-80 °C
Coolant Recirculator
-65 °C
-95 °C
-95 °C
Coolant Chiller, Coolant @ 10 °C, 0.75 l/min
-70 °C
-100 °C
-100 °C
Max Spectra per Second
88 (10 row crop mode), 81 (Full Vertical Bin)
System Window Type
Single UV-grade fused silica window, uncoated. Various AR coatings & MgF2 options available
Single UV-grade fused silica window; 1°wedge, AR coated on both sides, optimized at 900 nm
Blemish Specifications
Grade 1 as per sensor manufacturer definition
Features and Benefits
Features Benefits Peak QE up to 95%
Visible-optimized ‘BVF’ & infrared-optimized ‘BR-DD’ model
Fringe suppression technology as standard
Fringing greatly eliminated (Deep-Depletion) or minimized (Back-Thinned ‘BVF‘)
TE cooling down to -100°C
Critical for elimination of dark current detection limit - no inconvenience associated with LN2.
Ultravac™
Critical for sustained vacuum integrity and to maintain unequalled cooling and QE performance, year after year
Single UV-grade fused silica window
Best UV-NIR throughput performance, specific AR coating and wedge options available.
26 x 26 μm pixels
Optimized pixel format for high dynamic range and resolution.
Software-selectable pre-amplifier gain
Choice of best SNR performance of dynamic range at the touch of a button.
USB 2.0 connection
Ideal for laptop operation. Seamless operation alongside USB-based Shamrock spectrograph family
Compact and rugged platform
Ideal for OEM integration & laboratory setups.
Solis software for Spectroscopy
Comprehensive, user-friendly interface for simultaneous detector & spectrograph control.
Software Development Kit (SDK)
Ease of control integration into complex setups: Matlab, Labview, Visual Basic or C/C++.
Graphs
Graphs and Drawings
Dark Current
QE at 20 °C
QE at -100 °C
Dimensions - Front
Dimensions - Side
Connector Panel
Typical Setup
商品属性 [Active pixels ] 1024 x 127 [Pixel Size] 26 x 26 µm [Image area] 26.6 x 3.3 mm with 100% fill factor