1029 nm nanosecond laser module is a turnkey laser featuring high pulse energy and a pulse duration of less than 2 nanoseconds. High pulse energy and compact package make this laser suitable for Light Induced Breakdown Spectroscopy (LIBS) applications, as well as marking.
产品规格
Parameter Minimum Value Typical Value Maximum Value Central Wavelength, nm
1028
1030
1032
Longitudinal modes
-
multiple
-
Spectral line width FWHM, nm
-
0.7
1
Output power, mW
-
As per request
<500
Pulse duration, ns (polarized version)
1.4
1.5
1.8
Pulse duration, ns (non-polarized version)
1.1
1.2
1.5
Repetition rate, kHz (depending on pulse energy)
1.5
As per request
2.5
Pulse energy, µJ
-
As per request
100
Power stability, % (RMS, 8 hrs)
-
0.5 1
1
Pulse-to-pulse stability (RMS of peaks)
-
10
30
Transversal modes
-
TEM00
-
Beam Diameter at Aperture (1/e2), mm
-
1
-
Beam divergence (full angle), mrad
-
1
1.4
M2 effective
-
1.3
1.6
Jitter, μs
-
20
50
Jitter, %
-
3
8
Polarization direction
-
Horizontal (if polarized) 2
-
Polarization contrast
-
As per request
-
Control interface type
-
UART 0
-
Operation mode
-
APC (CW)
-
Input voltage, VDC
-
5
-
External power supply requirement
-
+5 V DC, 5A
-
Dimensions (L-W-H), mm
-
50 x 30 x 18 3
-
Beam height from the base, mm
-
10.4
-
Heat-sinking requirement, °C/W
-
0.5
-
Optimum heatsink temperature, °C
-
20
-
Warm up time, mins (cold start)
-
10
-
Temperature stabilization
-
Yes
-
Overheat protection
-
Yes
-
Storage temperature, °C (non-condensing)
-
-
-
Net weight, kg
-
0.29
-
Max. power consumption, W
-
25
-
Warranty, months (op. hrs)
-
14 (10000) 4
-
RoHS
-
Yes
-
CE compliance
-
- General Product Safety Directive (GPSD) 2001/95/EC
- (EMC) Directive 2004/108/EC-
Laser Safety Class
-
3B
-
OEM lasers are not compliant with
-
IEC60825-1:2014 (compliant using additional accessories)
-
1 The long term power test is carried out at constant laser body temperature (+/-0.1 °C) using an optical power meter with an input bandwidth of 10 Hz. The actual measurement rate has a period of about 20 seconds to 1 minute.
2 For lasers without integrated optical isolators.
0 Break-out-boxes AM-C8 and AM-H10 can be used for conversion of UART communication to either USB or RS232.
3 Excluding control interface pins and an output window/fiber assembly.
4 Whichever occurs first. The laser has an integrated operational hours counter.
Note: Product specifications are subject to change without prior notice to improve reliability, function or design or otherwise.
Typical spectrum
Typical spectrum of 1030 nm passive Q-Switch DPSS laser. Measured with 20 pm resolution.
Typical Near Field
Typical Far Field
Drawing
商品属性 [波长] 1030nm [脉宽] 1.5ns/1.2ns